Methods of
Hematology

shpslogo.jpg (6992 bytes)

Course Home Page

Study Aids

Glossary of Terms

 

6-2. MICROHEMATOCRIT

a. Principle. A capillary tube is filled with whole blood by capillary action to within 1 to 2 cm of the end. The unfilled end is sealed and the tube is centrifuged. After centrifugation, the capillary tube is placed in a reading device and the hematocrit value determined.

b. Procedure.

(1) If anticoagulated venous blood is the specimen, fill a plain capillary tube with blood. If blood without anticoagulant is used, fill a heparinized capillary tube with the blood specimen. A heparinized capillary tube is identified by red line on the tube specimen. A heparinized capillary tube is identified by a red line on the tube.

(2) Allow blood to enter two capillary tubes until they are approximately 2/3 filled with blood. (Air bubbles denote poor technique, but do not affect the results of the test.

(3) Seal the unfilled ends of the tubes with clay.

(4) Place the two hematocrit tubes in the radial grooves of the centrifuge head exactly opposite each other, with the sealed end away from the center of the centrifuge.

(5) Screw the flat centrifuge head cover in place.

(6) Centrifuge at 10,000 rpm for 5 minutes.

(7) Remove the hematocrit tubes as soon as the centrifuge has stopped spinning. Determine the hematocrit values with the aid of a microhematocrit reader. Results should agree within +2 percent. If they do not, repeat the procedure.

NOTE: Since there are a variety of readers available, it is necessary that the technician carefully follow the directions of the manufacturer for the particular device utilized.

c. Sources of Error.

(1) Inadequate mixing of the blood prior to sampling.

(2) Improper sealing of the capillary tube causes the blood to blow out of the capillary tube during centrifugation.

(3) Capillary tubes must be properly identified. Numbered holders for capillary tubes are available. Place the tubes in slots on the holder and record the numbers on the laboratory request slip.

(4) Improper centrifugation leads to varied results. For good quality control, maintain prescribed centrifuge speed and time.

(5) Misreading the red cell level by including the buffy coat causes elevated values.

d. Discussion.

(1) The microhematocrit technique is advantageous because of speed, and because only a small quantity of blood is necessary for the determination. An additional advantage is the ease with which this procedure is adapted to infants and small children. The microhematocrit technique requires only a simple capillary puncture whereas in the Wintrobe method venous blood oust be used. Another advantage is the use of disposable capillary tubes.

(2) If the microhematocrits cannot be read promptly, the capillary tubes must be properly identified and placed in a vertical position. Slanting of the cell layer will occur if tubes are left in a horizontal position for more than 30 minutes.

(3) Hematocrit results may also re obtained or computed through use of an electronic cell counter (Coulter Models). Results of computed hematocrits are generally 1.3 to 3.0 percent higher due to their method of derivation. An accurate erythrocyte count (three simultaneous electronic counts averaged) is multiplied by an accurate MCV measurement resulting in the area that the red cells would occupy if packed. The computation makes no allowance for the trapped plasma that always occurs to sate degree in manual packing procedures.

e. Normal Values.

(1) Birth: 62 percent.

(2) Age 1: 31-39 percent.

(3) Adult males: 42-52 percent.

(4) Adult females: 36-46 percent.


Curriculum design: David L. Heiserman
Publisher: SweetHaven Publishing Services

Copyright 2004 SweetHaven Publishing Services
All rights reserved